Description
SPECIFICATIONS:
The NEO-I is a premium, manually operated pneumatic infant resuscitator designed to deliver safe, precise, and controlled airway management for newborns and infants weighing under 10 kg. Fully compliant with the latest guidelines from ILCOR, AAP (NRP), and the AHA, the NEO-I is strongly recommended for the resuscitation of premature infants. By delivering operator-independent, consistent pressures, it helps successfully establish Functional Residual Capacity (FRC) while dramatically reducing clinical fatigue compared to traditional manual bag-mask squeezing.
Key Features
- Guidelines-Compliant Design: Engineered to align with strict global neonatal resuscitation protocols (AAP/NRP, ILCOR, and AHA), providing a safer and more reliable alternative to self-inflating manual resuscitation bags.
- Controlled PIP & Consistent PEEP: Delivers highly accurate and controllable Peak Inspiratory Pressure (PIP) along with reliable Positive End-Expiratory Pressure (PEEP). This helps improve lung volume, protect fragile lung tissue from barotrauma, and keep the alveoli open.
- 100% Pneumatic Operation: Purely gas-driven and requiring no electrical power supply, making it exceptionally reliable and continuously operational during critical power failures or emergency mobile patient transfers.
- Integrated Precision Manometer: Features an easy-to-read, fast-acting mechanical airway pressure gauge measuring from -10 to 80 cmH₂O (mbar) to enable non-stop visual tracking of delivery pressures.
- Adjustable Fail-Safe Pressure Relief (Pmax): Outfitted with a dedicated safety override dial to control the maximum pressure limit, ensuring pressures never accidentally exceed safe clinical thresholds (factory pre-set to a default of 40 cmH₂O).
- Ergonomic & Portable Layout: Built as a lightweight, compact system with a built-in top handle. It mounts seamlessly onto neonatal carts, walls, incubators, or radiant warmers across labor wards, NICUs, and transport teams.
Technical Specifications
- Target Patient Type: Neonates and infants weighing up to 10 kg (ideal for premature infants)
- Gas Flow Input Range: 5 L/min to 15 L/min
- Oxygen Concentration Control: 21% to 100% (dependent on your external gas supply or blender source)
- Manometer Gauge Scale: -10 to 80 cmH₂O (mbar)
- Max Pressure (Pmax) Range: 4 to 60 cmH₂O
-
PIP Performance Parameters (By Flow Rate):
- @ 5 L/min: 1 – 55 cmH₂O
- @ 8 L/min: 2 – 57 cmH₂O
- @ 10 L/min: 3 – 58 cmH₂O
- @ 15 L/min: 5 – 60 cmH₂O
-
PEEP Performance Parameters (By Flow Rate):
- @ 5 L/min: 0 – 8 cmH₂O
- @ 8 L/min: 1 – 17 cmH₂O
- @ 10 L/min: 1.5 – 23 cmH₂O
- @ 15 L/min: 3 – 26 cmH₂O